To Win:
Simply email your answer and your full contact information by December 1st to: Rebecca at rlsmith@pickeringlabs.com
The troubleshooting answer and winner congratulations will be published in the next issue.
Chromatography Quiz
Identify the error made when running the Carbamate chromatogram below and win a prize!
Pickering Standard: 1700-0063 Carbamate Test Mix, 2.5 µg/mL, inject 10 µL
Pickering Column: 1846150 Carbamate Column, C18, 4.6 x 150 mm
Normal Operating Conditions (for reference only, condition changes may be reflected in chromatogram):
Column Temperature: 42 °C
Flow rate: 1 mL/min
Eluent Gradient:

Post-column conditions for pesticide analysis:
Reagent 1: Hydrolysis reagent CB130
Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of CB910Reactor 1: 100 °C, 0.5 mL
Reactor 2: ambient. 0.1 mL
Reagent flow rates: 0.3 mL/min
Detection: Fluorometer ex 330 nm, em 465 nm