Chromatography Quiz #3
Identify the error made when running the Glyphosate chromatogram below and win a prize! Simply email your answer and your full contact information to Rebecca at rlsmith@pickeringlabs.com by July 1st in order to win. You will receive email confirmation when your submission is received, and the troubleshooting answer and winner congratulations will be published in the next issue (to be anonymous, please notify Rebecca in submission).

Glyphosate Analysis for US EPA Method 547
Normal Operating Conditions: (for reference only, changes may be reflected in chromatogram)

Customer’s Glyphosate Standard: 25 ppb, 100 µL injection

Pickering Column: 1954150 Cation-exchange Column for Glyphosate Analysis, 4 x 150 mm


Column Temperature: 55 °C
Flow rate: 0.4 mL/min

Eluent Gradient:









Post-column conditions for pesticide analysis:
Reagent 1: Oxidizing reagent – 100uL of 5% Sodium Hypochlorite in 950mL of GA116
Reagent 2: 100 mg of OPA, 2 g Thiofluor™ in 950 mL of GA104

Reactor 1: 36 °C, 0.5 mL

Reactor 2: ambient. 0.1 mL
Reagent flow rates: 0.3 mL/min

Detection: Fluorometer ex 330 nm, em 465 nm



Please Re-submit Your Answers to Amino Acid Quiz (No. 2)

Due to an unfortunate set of circumstances, all submitted answers to the previous Chromatography Quiz on Amino Acid Analysis have been lost. We sincerely apologize for the inconvenience to everyone who submitted an answer, and ask for all entries to be resubmitted. As we genuinely regret our technical difficulties, Rebecca will be sending prizes to all submissions, and grand prizes to everyone with the correct answer. Please resend your email, or resubmit your answer to rlsmith@pickeringlabs.com. Rebecca will send you confirmation via email that she has received your submission. All entries will be reviewed on June 11th, and prizes will be sent at that time. You can review the quiz at: http://pickeringlabs.blogspot.com/2010/01/chromatography-quiz-no-2.html. Again, our apologies, and if you have any questions please contact Wendy or Rebecca at 650-694-6700.